Set-Up Mode (cont.)
6. c-XX (“c” stands for Code Number)
(Selects The Attached Transducer Number)
The transducer number, full scale range and the units icon will show in the
window.
If the Window shows a range of “3333” (indicates an open cell which can be
programmed to the transducer) the associated transducer number has never been calibrated.
a. Press Enter to select this option.
b. Press Scroll-Up to select available transducer numbers (1-15).
c. When the number corresponds to the attached transducer, press the Enter key. The TM200 will now
use the appropriate torque calibration data for that transducer (if it has been previously calibrated).
7. CAL
(Calibration)
See the calibration section of this manual.
8. FIL
(Selects One of Three Electronic Low Pass Filters)
a. Press Enter to select this option.
The display shows the current filter setting:
Fil 1 = 500 (Hz)
Fil 2 = 1500 (Hz)
Fil 3 = 3600 (Hz)
b. Scroll to the desired filter setting and press enter to accept the new filter.
If you do not know what filter to use, set filter to 1500 Hz.